อุปกรณ์วัด Mini Controller (Resistivity Mini Controller)

BL983314 Resistivity Mini Controller

BL 983314

Resistivity Mini Controller

  • Easy to handle
  • Fire retardant casing
  • Selectable overdose protection system
  • Splash resistant cover

BL 983314 is a simple to operate resistivity controller designed for ultra pure water, reverse osmosis, and water conditioning applications. The BL 983314 resistivity controller is also ideal for continuous monitoring of process solutions. Set point and calibration are manually adjusted with a trimmer and the alarm relay allows for simple control.

Readings are automatically temperature compensated, with three different coefficients (β=2.4, 3.5 or 4.5 %/°C). The alarm contact can be used for connection to an alarm, pump, solenoid or dosing system.

The relay contact is open when readings are higher than the set point, while for measurements lower than set point, the relay contact is closed. The Hysteresis is typically 0.20 MW/cm from the set point.

Measurements are displayed on the LCD and the multicolored LED continuously indicates if the controller is in measurement, dosing or alarm mode. Users can choose automatic or manual dosing mode with a switch on the front panel. Manual control is particularly useful during maintenance operations, because it permits operators to enable or disable the dosing relay according to need.

The overtime control system advises users when the relay is active for too long, helping to prevent over-dosage.

Order Information:
BL 983314-0 (12 VDC) and BL 983314-1 ( 115/230V) are supplied with mounting brackets, transparent cover and instruction manual.

 Specification

Range   0.00 to 19.90 MΩ/cm
Resolution   0.10 MΩ/cm
Accuracy @ 25°C   ±2% F.S.
Probe   HI 3314 resistivity probe with 2 m (6.6’) cable (included)
Temperature Compensation   Automatic and linear, 5 to 50°C (41 to 122°F)
Temperature Coefficient   ß=2.4; 3.5; 4.5%/°C
selectable through jumper on the rear panel
Calibration   factory calibrated
Dosing Relay   maximum 2A (fuse protected), 250 Vac, 30 Vdc contact close when measure < set point
Setpoint   adjustable from 0 to 19.90 MΩ/cm
Overtime   adjustable, typically from 5 to approx. 30 minutes
Power Supply   See Chart
Dimensions   79 x 49 x 95 mm (3.1 x 1.9 x 3.7’’)
Weight   See Chart

Quotation and Ordering
131 Thailand Science Park. Innovation Cluster 1 Building, Klong 1, Klong Luang, Pathumthani 12120 Thailand.
Tel. +6681-847-6768, +662-191-3928 Fax +662-191-3929 Mr.Thoniwadh +6685-811-4779, Ms.Nareerat +6685-814-9653
Email: info@highents.com

สนใจขอใบเสนอราคาและสั่งซื้อสินค้า
131 อุทยานวิทยาศาสตร์ประเทศไทย อาคารกลุ่มนวัตกรรม 1 ต. คลองหนึ่ง อ. คลองหลวง จ. ปทุมธานี 12120
โทร. 081-847-6768, 02-191-3928 แฟกซ์ 02-191-3929 คุณโทณิวัฒน์ 085-811-4779, คุณนารีรัตน์ 085-814-9653
Email : info@highents.com

 

  •  

    >>> ขอใบเสนอราคาแบบออนไลน์ คลิก !!

  • Scroll To Top

    Contact Us